MSC calibration and autocorrelation data¶
MSC uses separate containers for main calibration, EUT calibration, and
autocorrelation. Every dataset is addressed by description.
Main calibration¶
Evaluate_MainCal creates processedData_MainCal[description]. All
non-scalar keys are fundamentally indexed by frequency:
PMaxRecandPAveRec: maximum and average received power by reference antenna position;PInputForEFieldandPInputForRecAnt: average input power;PInputVarationForEFieldandPInputVarationForRecAnt: maximum/minimum input-power ratio (the historical spellingVarationis contractual);EMax: maximum x/y/z component by probe position;EMaxT: maximum field magnitude;ACF: average chamber calibration factor;IL: average insertion loss;Enorm,EnormT,EnormAveXYZ,EnormAve, andEnormTAve: fields normalized bysqrt(PInput)and their averages;SigmaXYZ,Sigma24,SigmaXYZ_dB, andSigma24_dB: spread quantities.
EnormTmax2 remains as a historical, currently unpopulated frequency
container. Standard_Used names the selected evaluation standard.
CalculateLoading_MainCal creates an additional "empty,loaded" dataset
with Loading[frequency].
EUT calibration¶
Evaluate_EUTCal creates processedData_EUTCal[description] and within
it similarly creates PMaxRec, PAveRec, the four
PInput* tables, EMax, Enorm*, and Sigma*. It also adds:
CCFandCCF_from_PMaxRecChamber calibration factors from average and maximum received power.
CLFandCLF_from_PMaxRecChamber loading factors derived relative to the selected main calibration.
If field-probe values are absent, field-related tables are None at that
frequency while reference-antenna evaluation can still be available.
CalculateLoading_EUTCal additionally creates "empty,EUT" with
Loading[frequency].
Autocorrelation¶
rawData_AutoCorr[description] contains efield, tpos, and mg.
The field path follows the shared structure; tpos is the ordered list of
tuner positions actually measured.
Depending on the skip selection, Evaluate_AutoCorr writes to
processedData_AutoCorr[description]:
TunerPositions: sorted tuner positions;DistributuionOfr:n,r,pdf,cdf,rho, andrho0; the historical spelling is contractual;AutoCorrelation[frequency][probe_position][component]: cyclic autocorrelation coefficient sequence;NIndependentBoundarieswith the same index: estimated independent tuner positions;Statistic[frequency][probe_position][component]: samples, histogram, empirical and fitted distributions, and chi-square and KS p-values.independent_subset_usedandselected_sample_countshow whether an independent subset was used.
These statistical results depend on alpha, rho, rho0, lag,
every, and offset passed to evaluation. They are currently not
persisted as a dedicated metadata block and must be recorded separately for a
reproducible external reevaluation.
With explicit RC routing, raw sample bundles may additionally contain the
optional rx_capture field. Its schema, units and the 40 dB switching rule
are documented in MSC immunity and emission data. This applies to main
calibration, EUT calibration and autocorrelation; old bundles remain readable
without this field.