TEM immunity data

This page documents the persisted paths of the TEM/GTEM workflow. See Immunity results and EUT assessment for the domain meaning and assessment of EUT events.

A dataset is addressed by name:

raw = history.rawData_Immunity[description]
result = history.processedData_Immunity[description]

Raw data and run contract

Measurements use two parallel paths:

raw["power"][frequency][position][0][attempt]
raw["eutstatus"][frequency][position][0][attempt]

frequency is a numeric frequency value in Hz, position is the EUT orientation, and attempt indexes an initial or repeated attempt. The 0 level is the currently sole port index and readers must nevertheless include it.

A power entry contains value, pfwd, pbwd, testfield, am_enabled, and am_modulation. For historical reasons, value is again the mapping of measured quantities. pfwd and pbwd are power quantities and testfield is a field-strength quantity. leveling_* keys are added according to the leveling strategy.

An eutstatus entry contains at least value, pfwd, and pbwd. The current workflow adds testfield, am_enabled, am_modulation, and eut_event, plus optional leveling_* keys. value is the legacy-compatible EUT status; eut_event is the structured event.

exposure_status[frequency][position] has status, reason, detail, and testfield. Relevant status values are completed, retry_pending, not_evaluated, observation_pending, and not_executed. The last value is stored, for example, when amplifier protection is triggered. monitor_diagnostics is a list of EUT-monitoring error and status records.

Important workflow and provenance fields

Key

Meaning

requested_frequencies, requested_positions

Planned scope used for completeness checking.

reference_selection, reference_summary

Selected TEM verification and summary of its suitability.

immunity_target_efield

Required field strength as a quantity.

immunity_reference_power_selection

Selected or interpolated reference power for each frequency.

am_headroom_policy, am_modulation

Protection/evidence policy and AM parameters actually configured.

position_coverage

Validation of required EUT orientations.

performance_criterion, eut_event_policy

Assessment criterion and actions for monitor events.

post_exposure_timeout

Observation duration after the disturbance is switched off.

measurement_graph_mode

Logical graph context used, normally "GTEM".

immunity_run_contract

Stable snapshot of settings that must remain unchanged when resuming from autosave.

mg

Persisted measurement-graph description, not an open device connection.

immunity_run_contract is an independently versioned subdocument. Its current schema_version is 2. It stores quantities as mappings with __quantity__, value, uncertainty, and unit so settings can be compared stably. This field does not replace SCUQ quantities in measurement data.

Processed data

The primary result path is:

result["result"][frequency][position][attempt]

Each attempt contains eutstatus, eut_event, passed, effective, testfield, pfwd, pbwd, am_enabled, and am_modulation. leveling is optional. Only the last attempt for a position has effective == True; older attempts remain for traceability. Under criterion A, an earlier impairment can still affect the overall assessment after a later passing attempt.

Top-level keys of processedData_Immunity[description]

Key

Meaning

result

Evaluated attempts by frequency and position.

performance_criterion, performance_assessment

Criterion A/B/C and structured assessment for each measurement point.

overall_passed, overall_status, complete

Overall decision and completeness. overall_status is passed, failed, or incomplete.

incomplete_exposures

Missing or unevaluable requested exposures.

leveling_summary

Leveling metadata by frequency and position.

monitor_diagnostics

Copy of diagnostics from raw data.

reference_selection, reference_summary

Reference used and its suitability summary.

immunity_target_efield, immunity_reference_power_selection

Target field and power reference used.

am_headroom_policy, am_modulation

AM evidence and modulation.

eut_event_policy, position_coverage, post_exposure_timeout

Assessment and workflow context.

An incomplete_exposures entry has frequency, position, status, reason, detail, and testfield. complete is True only when every requested frequency-and-position combination has a valid EUT status. Consequently, do not interpret overall_passed == False without overall_status: it can mean a technical failure or an incomplete run.